ความไว
0.53 ng/mL
ช่วงการตรวจวัด
1.56-100 ng/mL
ชนิดตัวอย่าง
Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids,

ELISA

Product Details

Product Name VD2(Vitamin D2) ELISA Kit Species Universal
Uniprot ID N/A Alternative Names Ergocalciferol; Deltalin; Drisdol; Calcidol
Detection method Competitive Sensitivity 0.53 ng/mL
Standard 100 ng/mL Detection Range 1.56-100 ng/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids, Sample volume: 50μL
Reaction time 1.5H Research Area Metabolic pathway;Neuro science;Nutrition metabolism;Bone metabolism;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with VD2. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to VD2. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of VD2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD
100.00 0.278
50.00 0.378
25.00 0.572
12.50 0.833
6.25 1.128
3.13 1.523
1.56 1.837
0.00 2.501
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 1.970 4.200 28.090 2.050 3.140 34.970
Standard deviation 0.210 0.460 1.450 0.260 0.580 1.810
C V (%) 4.21 5.32 2.50 3.47 7.47 5.62
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 92-106 99
EDTA plasma (n=8) 88-103 95
Cell culture media (n=8) 80-95 86
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 87-103 85-95 87-101
Average (%) 92 90 93
1:4 Range (%) 97-105 85-95 88-100
Average (%) 105 90 93
1:8 Range (%) 80-90 85-96 95-103
Average (%) 87 86 96