Human ELISA Kits

Human IgG2(Immunoglobulin G2) ELISA Kit

Catalog #: RE3190H

Manufacturer: Reed Biotech

Size: 96 T48 T

Price: on inquiry

Reactivity
Human
Sensitivity
0.94 μg/mL
Assay Range
1.56-100μg/mL
Sample Type
Serum, Plasma, Tissue homogenate and Other biological samples;

ELISA

Product Details

Product Name Human IgG2(Immunoglobulin G2) ELISA Kit Species Human
Uniprot ID P01859 Alternative Names IGHG2
Detection method Competitive Sensitivity 0.94 μg/mL
Standard 100μg/mL Detection Range 1.56-100μg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area Metabolic pathway;Cardiovascular biology;Hepatology;Nutrition metabolism;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human IgG2. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to Human IgG2. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human IgG2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(μg/mL) OD
100.00 0.257
50.00 0.357
25.00 0.551
12.50 0.812
6.25 1.107
3.13 1.502
1.56 1.816
0.00 2.480
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(μg/mL) 2.34 15.36 25.43 5.28 25.63 30.16
Standard deviation 0.27 0.65 1.77 0.06 0.94 2.61
C V (%) 3.24 6.96 6.21 2.71 4.90 4.37
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 82-94 88
EDTA plasma (n=8) 97-107 102
Cell culture media (n=8) 87-99 93
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 87-95 89-103 85-97
Average (%) 91 97 86
1:4 Range (%) 80-91 89-103 95-102
Average (%) 87 95 97
1:8 Range (%) 97-103 80-92 97-105
Average (%) 99 91 104
1:16 Range (%) 95-101 88-97 87-94
Average (%) 101 93 91

Assay Procedures

View this product on the manufacturer website (Reed Biotech) ↗

Request a quotation

Fill in the form and the Envision Sciplus team will reply with a quotation and delivery time.

Product Human IgG2(Immunoglobulin G2) ELISA Kit Catalog #: RE3190H
Sent to info@envisionsci.com